ORP Meter 3 in 1 pH ORP TEMP Tester pH Meter with Backlight Multi-parameter Digital Tri-Meter Water Quality Monitor
Specifications:
Measuring range:
PH:0.00-14.00 pH ; ORP:-1999-1999mV ; Temp:0-50℃
Resolution:
PH:0.01pH ; ORP:1mV ; Temp:0.1℃
Accuracy:
PH:±0.1pH ; ORP:±5mV ; Temp:±1℃
Automatic Temperature Compensation:0℃-50℃
Battery Type:4×1.5V(AG13)(not include)
operating Temperature:0℃-50℃
Calibration:Manual,at one point through trimmer
Dimensions:188mm×35mm×35mm
Weight:156g
OPERATION:
1.Remove the protective cap.Turn on the meter.
2.First rinse the electrode with distilled water,and suck it with filter-paper.
3.Immerse the electrode in the solution to be tested.
4.Stir gently and wait for the reading to stabilize.Do not tap it against the glass.
5.Press the"pH/ORP"button,choice the PH or ORP for measuring.
6.Press the "C/F"button to switch between Celsius and Fahrenheit.
7.After use,turn off the meter.Rinse the electrode with distilled water to minimize contamination.
8.Always replace the protective cap after use.
Calibration:
1)PH
1.Rinse the ph electrode first with distilled water,and suck it with filter paper.
2.Turn on the meter. Press the "pH/ORP"button,choice the PH for measuring.
3.Immerse the ph electrode in pH6.86(or pH7.00)buffer solution,gently shake the meter and wait for a stable display.
4.Regulate the trimmer with the screwdriver provided until display shows "6.86"("7.00")at buffer solution temperature of 25℃.
5.Rinse the ph electrode with distilled water,and suck it with filter paper (not included)
6.Immerse the pH electrode in pH4.01 or pH9.18(pH10.0)buffer solution,gently shake the electrode and waiting for the reading to stabilize.
The showing value and buffer solution's ph value is within error range.
7.The calibration of the instrument's pH range is now complete.
Important:
The ph meter should be recalibrated under following situations:
-
The meter has been used for long time since the last calibration or use.
-
The meter has been used in particularly continual conditions.
-
The utmost accuracy is required.
2)ORP
1.Turn on the meter.Press the"pH/ORP" button,choice the ORP for measuring.
2.Rinse the electrode with distilled water.
3Immerse the electrode into the fresh 256mV ORP standard solution,then stir
it gently,
4)Allow the reading to stabilize and with a small screwdriver turn the calibration
trimmer until the display shows"256".
5)Rinse the electrode with distilled water and suck it with filter paper.
6)Switch off the meter.Replace the battery compartment cap.
The calibration of the instrument's ORP range is now complete.
ELECTRODE REPLACEMENT:
1.Remove the protective cap.
2.Unscrew the plastic ring on the top of the electrode.
3.Pull out the electrode and replace it with a new one.
Make sure the gaskets are in place before screwing back the ring.